Rabu, 06 April 2016

IKA Colorimeter C 6000 global standards Package 2/10

IKA Colorimeter C 6000 global 

standards Package 2/10


Description 

The C 6000 global standards oxygen bomb calorimeter combines modern technology, variability and automation (adiabatic, isoperibol; dynamic modes) in one instrument. It operates according to all bomb calorimeter standards, such as e.g. the DIN, ISO, ASTM, GOST and GB. The operator can choose between three different starting temperatures (22 ºC, 25 ºC, 30 ºC) in each measuring mode: adiabatic, isoperibol or dynamic. Due to the spherically shaped decomposition vessel head, the wall thickness was able to be reduced, allowing for faster heat transfer which results in a shorter time measurement. Due to the variety of different interfaces (PC, Ethernet, SD-card, balance, printer) this unit is easily adaptable depending on the customer’s specific application needs. Further adaption to data management and LIMS is possible with our calorimeter software C 6040 Calwin (Accessory). 

- Automatic ignition 
- Automatic water filling and draining 
- Automatic oxygen filling, venting and flushing 
- RFID technology for automatic decomposition vessel identification 
- New design of the decomposition vessel allows for easier and faster sample preparation 
- Can be operated with a chiller (e.g. KV 600) 
- Easy and convenient touch screen operation 
- Control chart view and correction calculation of globally used standards 
- Ethernet interface for data management via FTP server or connection of a network printer 
- Removable SD-card allows for easier data measurement management, in addition to software updates 

The C 6000 global standards Package 2/10 consists of: 
- C 6000 global standards (0003780000) 
- C 6010 Decomposition vessel, standard (0003770000) 


Technical Data 

Measuring range max.40000 J
Measuring mode adiabatic 22°Cyes
Measuring mode dynamic 22°Cyes
Measuring mode isoperibol 22°Cyes
Measuring mode adiabatic 25°Cyes
Measuring mode dynamic 25°Cyes
Measuring mode isoperibol 25°Cyes
Measuring mode adiabatic 30°Cyes
Measuring mode dynamic 30°Cyes
Measuring mode isoperibol 30°Cyes
Measurements/h adiabatic5
Measurements/h dynamic6
Measurements/h isoperibol4
Reproducibility adiabatic (1g benzoic acid NBS39i)0.05 %RSD
Reproducibility dynamic (1g benzoic acid NBS39i)0.15 %RSD
Reproducibility isoperibol (1g benzoic acid NBS39i)0.05 %RSD
Touchscreenyes
Working temperature min.22 °C
Working temperature max.30 °C
Temperature measurement resolution0.0001 K
Cooling medium temperature min.12 °C
Cooling medium temperature max.27 °C
Cooling medium permissible operating pressure1.5 bar
Cooling mediumtap water
Type of coolingflow
Flow rate min.60 l/h
Flow rate max.70 l/h
Rec. flow rate at 18°C60 l/h
Oxygen operating pressure max.40 bar
Interface scaleRS232
Interface printerUSB
Interface PCRS232
Interface test rackyes
Interface ext. keyboardyes
Oxygen fillingyes
Degasificationyes
Decomposition detectionyes
Decomposition vessel C 6010yes
Decomposition vessel C 6012no
Analysis according to DIN 51900yes
Analysis according to DIN EN ISO 1716yes
Analysis according to DIN EN ISO 9831yes
Analysis according to DIN EN 15170yes
Analysis according to DIN CEN TS 14918yes
Analysis according to ASTM D240yes
Analysis according to ASTM D4809yes
Analysis according to ASTM D5865yes
Analysis according to ASTM E711yes
Analysis according to ISO 1928yes
Analysis according to GB T213yes
Dimensions (W x H x D)500 x 425 x 450 mm
Weight29 kg
Permissible ambient temperature20 - 25 °C
Permissible relative humidity80 %
Protection class according to DIN EN 60529IP 20
RS 232 interfaceyes
USB interfaceyes
Voltage220 - 240 / 100 - 120 V
Frequency50/60 Hz
Power input1700 W