Rabu, 06 April 2016

IKA Colorimeter C 7000 basic equipment set 1

IKA Colorimeter C 7000 basic 

equipment set 1


Description 

The C 7000 is the first IKA® calorimeter with a completely dry system for measuring the gross calorific value of solid and liquid samples. The temperature is measured directly in the decomposition system. This results in measurement times in the range of 3 to 7 minutes (depending on the sample). The system can manage up to 8 different decomposition vessels using a code ring scheme. 
- High sample frequency 
- Precise and reproducible determination of gross calorific values according to ISO 1928 
- Reduction of routine work through automatic application flow 
- Automatic decomposition vessel identification 
- Interface connections for scale, printer and PC 
- User-friendly software C 5040 CalWin for controlling the calorimeter and administration of measuring data 
- The decomposition vessel can be changed over to use combustible crucibles C 14 
C 7000 basic equipment set 1 consisting of: 
C 7000 Measurement cell 
C 7010 Decomposition vessel, standard 
C 7002 Cooling system 
C 48 Oxygen station 


Technical Data 

Measuring range max.30000 J
Measuring mode double dry (ISO 1928)yes
Measuring time double dry approx.3 min
Double dry reproducibility (1g benzoic acid NBS39i)0.2 %RSD
Working temperature max.30 °C
Temperature measurement resolution0.0001 K
Cooling medium temperature min.12 °C
Cooling medium temperature max.30 °C
Cooling medium permissible operating pressure9 bar
Cooling mediumtap water
Type of coolingflow
Flow rate min.120 l/h
Flow rate max.180 l/h
Oxygen operating pressure max.40 bar
Interface scaleRS232
Interface printerCentronix
Interface PCRS232
Oxygen fillingno
Degasificationno
Decomposition detectionyes
Decomposition vessel C 7010yes
Dimensions (W x H x D)310 x 395 x 490 mm
Weight43 kg
Permissible ambient temperature18 - 30 °C
Permissible relative humidity80 %
RS 232 interfaceyes
Voltage220 - 240 V
Frequency50/60 Hz
Power input260 W

Tidak ada komentar: