SK- SATO Infrared Thermometer Model SK-8300 Range:-30 to 1550℃ Cat.No.8268-00


SK- SATO Infrared Thermometer Model SK-8300 Range:-30 to 1550℃ Cat.No.8268-00


With the wide measuring range of -30 to 1550℃, this thermometer can measure extremely high temperature to be used for wider applications. Its "distance to spot ratio", 50:1,enables to measure smaller spots. Emissivity is adjustable from 0.10 to 1.00 in 0.01 steps.
Features

OInstance surface temperature measurement:
The advantage of the non-contact infrared method is quick-response measurement. And this single-hand-held unit allows easy operation by use of the trigger type measurement button.
OHigh-temperature measurement:
Wide measurement range of -30 to 1550°C The laser marker indicates the spot of measurement with a clear point.
ODistance to measurement spot ration (D:S):
About 50:1 (D:Measurement distance, S:Spot)
OAlarm function of upper and lower thresholds:
Alarm is triggered when a measured temperature value exceeds upper or lower set value.
OData Memory function:
Store up to 10 data
OVarious indications:
MAX(Maximum temperature value), MIN(Minimum temperature value),AVG(Average temperature value), ΔT(Difference between maximum and minimum temperature values)
OAuto Measurement function:
Continuous measurement without holding the trigger down
OBacklight function:
The backlit LCD facilitates to read measured values in dark.
OAuto Power-Off function:
This function conserves battery power when you forgot to turn the power off.
OTripod fixable:
The unit can be fixed to a tripod for measurement.
OClass II Laser Product with PSC mark:
This product is designed to comply with portable laser instrument Consumer Product Safety Act.

Specifications

Cat. No.No.8268-00
ModelSK-8300
Measuring range-30°C to 1550°C
Accuracy-30.0°C to -20.1°C : ± 3°C
-20.0°C to 100.0°C  : ± 2°C
Other then above : ± 2%rdg
(Condition: Ambient temperature at 23°C ± 5°C, calibrated by black body emissivity at 0.95
Resolution0.1°C
EmissivityAdjustable at 0.01 within 0.10 to 1.00
Refer to the emissivity table
Distance to spot ratioD:S = approx. 50:1 (D: Distance, S: Spot size)
Sensing elementThermopile
Spectral response8 to 14μm
Laser MarkerClass II laser product with PSC mark
Wavelength: 650mm Output: less than 1mW
Operation ambient0 to 50°C for temperature, less than 85%rh for humidity (no condensing)
Storage ambient-10 to 50°C for temperature, less than 85%rh for humidity (condensing)
Power requirements9VDC (Battery (006P) x 1 pc.)
Power consumption40mA (When laser and backlight ON)
Battery lifeApprox. 8 hours continuous use (use of alkali battery)
(When laser and backlight ON under normal temperature)
BacklightLCD with backlight
Weight and dimensionsApprox 280g (including battery) (W)51 x (H)200 x (D)166mm
Accessories9V alkali battery x 2 pcs. (one is factory set), hand strap, hard case

*All specifications subject to change without notice

Distance to spot ratio (D:S)
The SK-8300 focuses energy from an object onto its detector at the rate of approximately 50:1 (D = Distance, S = Spot size). The SK-8300 will read an area of 19mm in diameter from 500mm away. For higher accuracy, make sure the area of the object at least twice as large as the spot being measured.
For better measurement:
Square: measured object Circle: measurable range:
GoodWithin the range, but for better measurement, get near to the object.It measures ones in addition to object. Get near to the measurable range.
*Caution
It is dangerous if the laser beam gets into an eye. Do not look directly into the laser emitting window and do not point the laser at people.
(Class II) A visible light of a wavelengths in the visible region, 400nm through 700nm, with an output lever, approximately 1mW or less, that is normally regarded harmless with respect to humans' physical defense capabilities.

Caution for safety 
Read this manual thoroughly to use SK-8300 safety
Do not point the laser at people Class II laser product: MAX1.0mW 650nm