Selasa, 29 Maret 2016

SK - SATO WBGT Heat Stress Monitor Model SK-150GT Cat.No.8310-00


SK - SATO WBGT Heat Stress Monitor Model SK-150GT  Cat.No.8310-00





The SK-150GT is a handy type device that can measure temperature, humidity and globe temperature to approximate the WBGT index (*). 
This can be used as a guide for preventing heat related illness and estimating heat stress of working and activity related environments.


PURPOSE OF USE
The SK-150GT is a handy type instrument that can measure WBGT (*) index, temperature, humidity and globe temperature. This can be used as a guide for preventing a heat illness and estimating heat stress of working and activity related environments.
 *WBGT: Wet Bulb Globe Temperature
By visually observing the sensory temperature, a heat stress can be prevented. The SK-150GT would come in handy at following situations.
-
Gymnastic classes or outdoor activities in schools
-
Indoor and outdoor sports activities
-
In high temperature working environments
-
Construction sites
-
Farms
* WBGT Index :
The process for determining the WBGT (Wet Bulb Globe Temperature) is described in ISO 7243 / JIS Z8504, Hot Environments - Estimation of the Heat Stress on Working Man, based on the WBGT Index. The WBGT index is a sensory temperature which can be calculated from wet bulb temperature, globe temperature (radiation heat) and dry bulb temperature. It is necessary to set the measuring ambient before starting since different calculation is needed depending on its measuring ambient.
(1)Indoor (IN): when measuring indoor or outdoor away from direct sun light.
WBGT Index(°C) = 0.7 x Wet Bulb Temperature(°C) + 0.3 Globe x Temperature(°C)
(2)Outdoor (OUT): when measuring outdoor, under direct sun light.
WBGT Index(°C) = 0.7 x Wet Bulb Temperature(°C) + 0.2 x Globe Temperature(°C) + 0.1 × Dry Bulb Temperature(°C)
Heat Stress :
Heat stress is a condition that temperature regulation function decreases by high temperature ambient. This could cause a heat related illness and even ultimately death. The risk can be reduced by observing the surrounding ambient and by being aware of precautions of heat stock such as overheating and dehydration.
FEATURES
  • Alarm: The alarm is triggered when the measured WBGT value reaches the set value.
  • HOLD function, MAX/MIN function
  • Measuring ambient of WBGT index can be set.
  • Auto power-off (in 20 min.)
  • Carrying case for damage prevention and storing of the instrument.
No. 8310-60 Tripod for SK-150GT Model EX-430SK
Height:1513mm   Weight: 1.2kg

A tripod is recommended for continuous measurement. Select a location free from effects of radiation heat.
SPECIFICATIONS
ModelSK-150GT
Measuring rangeAir Temperature (TA)0.0 to 50.0°C
Globe Temperature (TG)0.0 to 80.0°C
WBGT Index0.0 to 50.0°C
Humidity10.0 to 95.0%rh
AccuracyAir Temperature (TA)±0.6°C
Globe Temperature (TG)±2°C
WBGT Index±2°C
Humidity±3% rh (20 to 90%rh at20 to 30°C)
±5% rh (other than above)
ResolutionAir Temperature (TA)0.1°C
Globe Temperature (TG)0.1°C
WBGT Index0.1°C
Humidity0.1% rh
Sampling timeapprox. 2 seconds
WBGT alarmSetting range20.0 to 49.9 (no cancelation)
FunctionsChange-over switch for WBGT Index of indoor and outdoor
Auto power-off in approx. 20 min.
MAX/MIN
HOLD
Operation ambient0 to 50°C, below 95%rh (no condensing)
Storage ambient-10 to 50°C (no condensing)
Power requirement9VDC Battery 6F22 (006P) x 1
Battery lifeapprox. 600 hours in continuous use (without use of alarm)
DimensionsMain Unit : (W) 71 x (H) 27.5 x (D) 30 mm (without globe)
Globe : Ø 60 mm
Weightapprox. 240 g (with battery)
Standard accessories9VDC Battery 6F22 (006P) x 1
Carrying case